Description

These ATI aerosol generators run on compressed air, and are ideal for the DOP leak testing of cleanroom ceiling systems, clean enclosures, mini environments, workstations, and many other clean applications.  The 4B can support air flows up to 8,100 CFM, while the 4Blite is ideal for smaller applications, handling up to 5,000 CFM.  The optional hose adapter allows for introducing the aerosol into positive pressure systems using a ¾ inch ID hose.


Details

TDA-4B – 6 laskin nozzles, 50-8,100 CFM @ 20psig

TDA-4Blite – 3 laskin nozzles, 50-5,000 CFM @ 20psig

Stainless Steel body with large fill reservoir



 

ATI TDA-4B/4B Lite

Specifications


Aerosol Concentration:

TDA-4B: 10 mg/l @ 8,100 CFM @ 20psig

TDA-4Blite: 10 mg/l @ 4,050 CFM@ 20psig


Dimensions:

TDA-4B: 10 x 11 x 9, 16 lbs

TDA-4Blite: 10 x 8 x 9, 12 lbs


Power:

None required (compressed air operated)


Consumables:

DOP/PAO/Emory 3004